F5100 : High Precision Coriolis Mass Flowmeter

Model : F5100
Download :

Description

Special Features :
b Apply high stability and patented technology platinum RTD sensor
b Uses proprietary technology “balanced structure package”, medium temperature self-compensation.
b Proprietary algorithms; can achieve high linearity, high repeatability, high precision.
b Range ratio 1000:1, also can be expanded according to user requirements.
b Enables measurement of small flow large diameter,minimum flow rate can be measured as low as zero, resolution0.001m/s.
b No moving parts, vibration can be neglected.
b Straight pipe less demanding 1-2D
b Unrelated with the medium temperature and pressure.
b Proprietary high-temperature method, medium temperature up to +510 .
b Humidity algorithm used to achieve high precision measurement
b Flow signal enter 12-point dynamic correction, built-in 10-point amendment.
b Dynamic on-line correction current / voltage output.
b On-line continous flow,easy handling and maintenance

 

 

Additional information

Measuring range

: 0-120Nm/s (20C,101.33Kpa)

Accuracy

: ± 0.1% of reading ± 0.5% full scale

Range ratio

: usually 100:1 (depending on the calibration of flow range)

Diameter range

: 6000mm

Application scope

suitable for all types of single or mixed gases, containing dust, sand, moisture and various corrosive gases.

Environmental temperature range

-40 – +85 (no display); -30 – +70 (showing); humidity
is less than 90% RH.

Medium temperature range

: -40 – +100; -40 – +200; -40 – +450; -40 – +510

Sensor Material

: 316 stainless steel, Hastelloy, Titanium

Probe rod material (protection cover)

: 316 stainless steel (standard), Hastelloy

Analog Output

: Flow:4-20mADC,temperature:4-20mADC,maximum
load:10000

Communication

: Serial output RS232/RS485

Power supply

: 24VDC/600mA

Alarm

: 1-2-way relay output, 5A/220V, 5A/30VDC, type settings

Large-screen LCD Display

: 7 instantaneous flow, 8 cumulative flow rate

Process Pressure

: 1.6Mpa (maximum 20MPa)

Installation Process Type

: Plug-in (card sets, card sets + ball valve, flange
connection), pipeline type (flange, screw connection)

Protection class

: IP66